Journal: Journal of Applied Clinical Medical Physics
Article Title: Dose rate versus gantry speed performance evaluation for slow gantry speeds using DICOM RT plans
doi: 10.1002/acm2.13786
Figure Lengend Snippet: Distribution of monitor unit (MU) per degree for stereotactic body radiation therapy (SBRT) treatment plans. The dashed yellow line shows the “critical” MU per degree value (1.67 MU/deg) where, once exceeded, the dose rate is maximized and the gantry speed is modulated. The dashed red line on the histogram represents the maximum MU per degree value tested using the vendor‐provided DRGS DICOM RT Plan file. The dashed green line indicates maximum MU per degree values in the modified DRGS DICOM RT Plan file.
Article Snippet: The current range of MU per degree values that Varian's DRGS DICOM RT Plan file tests was first determined.
Techniques: Modification